Category
- Material Characterization » Thermal Characterisation
Booking Details
Measurement Method:-Heat Flux
Temperature Range:-(-150C to 725C)
Scanning Rate:-0.1 to 100C
Gas Type:-N2,Air,Ar,O2
Gas Flow Rates:-50 to 500 ml/m
Facility Management Team and Location
office.saif@iitb.ac.in
022- 2576-7691
nmagadi@iitb.ac.in
022- 2159-6889
Facility Features, Working Principle and Specifications
Facility Description
Heat Flux DSC
A technique in which the sample and reference crucible are heated uniformly (under the same conditions) with a definite amount of heat. The difference in temperature between the sample and the reference material gives the heat capacity of the sample and is measured as a function of time and temperature.
DSC is a thermal analytical technique that looks at how a material’s heat capacity (Cp) is changed by temperature. It measures the amount of heat released or absorbed by a material and hence DSC enables the measurements of the transition such as the glass transition, melting, and crystallization.
Instructions for Registration, Sample Preparation, User Instructions and Precautionary Measures
- Only online registration through the SAIF webpage will be accepted.
- Slots will be provided as per the queue. If the appointment is given but the user cannot come, a mail should be immediately sent to [at] iitb [dot] ac [dot] in to cancel his/her slot.
- Data will be sent to the registered mail ID. USB drives are not allowed due to computer virus problems.
- MSDS form has to be filled and uploaded.
- External users have to upload proof of payment.
- At least 30-40 mg of sample should be submitted.
- Sample size should be < 5mm in diameter and < 2.5mm in thickness.
- Samples containing halogens should not be sent and will not be accepted for analysis.
- Samples should not be toxic or hazardous.
- Heating program should be less than the decomposition temperature.
Samples containing corrosive substances that will react with Aluminium/Platinum/Ceramic pans should not be sent for analysis by DSC.
Charges (including 18% GST) *For Thermal Analysis, standard ramp is 10 ℃/min. Heating rates less than standard rate will be charged by hours.
|
Charges for Analytical Services in Different Categories
Applications
Glass Transition temperature
Melting and Crystallization temperature
Percent Crystallinities
Compositional analysis of blends
Thermal histories
Thermal denaturation for proteins
Purity Determination
Polymorphism
Drug Compatibility studies, etc.